accordance with
IEC 61000-4-8
the typical val-
ues as they are
in a
commercial and
clinical environ-
ment.
REMARK : U
r
is the alternating current in the network before the application
of the test level.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Tests of Resist-
ance to Dis-
ruption
IEC 60601-1-2:2014
Test Level
Level of
Agree-
ment
Electromagnetic En-
vironment
Use portable and mo-
bile radio sets at a
safe distance from the
device (including the
leads) not closer than
the recommended
protective distance,
which is calculated
according to the equa-
tion suitable for the
RF emission frequen-
cy.
Recommended pro-
tective distance:
Conducted
high frequency
disturbance
variables in ac-
cordance with
IEC 61000-4-6
6 V/m in ISM fre-
quency bands be-
tween 150 kHz and
80 MHz
6 V/m in
ISM fre-
quency
bands be-
tween 150
kHz and 80
MHz
d
= 1.2
Radiated high
frequency dis-
turbance varia-
bles in accord-
ance with IEC
61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to
2.7 GHz
d
= 1.2
80 MHz
to 800 MHz
DRY 1.4, DRY 2.4 | Remarks for HF-emission and immunity |
117
3631B EN 20210601 1655
Summary of Contents for 5366/500
Page 1: ...DRY 1 4 DRY 2 4 5366 510 5366 500 User Manual 3631B EN 20210601 1655 ...
Page 12: ...12 DRY 1 4 DRY 2 4 DRY 1 4 DRY 2 4 3631B EN 20210601 1655 ...
Page 57: ... The display DRY 1 4 DRY 2 4 Introduction 57 3631B EN 20210601 1655 ...
Page 141: ...DRY 1 4 DRY 2 4 Plug Play Installation Manual 141 3631B EN 20210601 1655 ...