For Technicians | AfiPass II Installation Guide
8.3
Calibrating the Antennas
Calibrate each antenna for optimal resonance.
To calibrate the antennas:
1.
Set the technician's tool to CAL Mode. On the Technician toolbox:
a.
Turn ON TS1.
b.
Turn OFF TS2.
Indication: The display shows 'CAL T'.
2.
Set the antenna number:
Press PB1 to increase in increment by 1.
Press PB2 to decrease in increments of 1.
Indication: The display alternates between the antenna number and the calibration value
(calibration values range from 1-127).
3.
Calibrate the selected antenna: Press PB1 and PB2 simultaneously. Hold these down for
3 seconds. The calibration process takes about 30 seconds.
Indication: The display shows 'CL**XX'
Indication: Once the Calibration process is complete, if successful, the display shows 'CL OK'
and a sound is emitted for 4 seconds.