Aesculap ELAN 4 Instructions For Use/Technical Description Download Page 24

21

USA

9.2

Fuse replacement

 

DANGER

Risk of death by electric shock!

Unplug the device before changing the fuses!

Specified fuse set: 2 units IEC 127 – T 6.3 A switching capacity H (1 500 A 
at 250 V/50 Hz)

Use a small screwdriver to release the clip on the fuse holder 

12

.

Pull out the fuse holder 

12

.

Replace both fuses.

Reinsert the fuse holder 

12

 in such a way that it audibly snaps into 

place.

Note

If the fuses regularly blow, the unit is defective and has to be repaired. see 
Chapter 10.

10. Technical Service

 

DANGER

Danger to life of patients and users in case of malfunctions and/or 
failure of protective measures!

Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.

 

CAUTION

Modifications to medical equipment may result in loss of guaran-
tee/warranty claims and any approvals.

Do not modify the product.

For service and repairs, please contact your national 
B. Braun/Aesculap representative.

Modifikationen an medizintechnischer Ausrüstung können zu einem Ver-
lust der Garantie-/Gewährleistungsansprüche sowie eventueller Zulassun-
gen führen.

Für Service und Instandsetzung wenden Sie sich an Ihre nationale 
B. Braun/Aesculap-Vertretung.

Service addresses

Aesculap Technischer Service

Am Aesculap-Platz 

78532 Tuttlingen / Germany

Phone:

+49 (7461) 95 -1601

Fax:

+49 (7461) 14 -939

E-Mail:

[email protected]

Or in the US: 

Aesculap Inc.

Attn. Aesculap Technical Services 

615 Lambert Pointe Drive 

Hazelwood 

MO, 63042 USA 

Aesculap Repair Hotline 

Phone:

+1 (800) 214 -3392 

Fax:

+1 (314) 895 -4420

Other service addresses can be obtained from the address indicated above.

11. Accessories/Spare parts

11.1

ELAN 4 electro motor cables, applied parts and 
foot controls

Art. no.

Designation

GA804

ELAN 4 electro motor cable with hand lever
(from software version 2.00)

GA805

ELAN 4 electro motor cable with hand control
(from software version 2.00)

GA806

ELAN 4 electro motor cable for foot control

GA808

ELAN 4 electro foot control

GA810

ELAN 4 electro wireless foot control (from soft-
ware version 3.00)

GA822

ELAN 4 electro perforator driver

GA824

ELAN 4 electro Lowspeed motor with intra cou-
pling

GA831

ELAN 4 electro sagittal saw

GA832

ELAN 4 electro reciprocating saw

GA833

ELAN 4 electro transverse saw

GA836

ELAN 4 electro micro sagittal saw

GA837

ELAN 4 electro micro reciprocating saw

GA844

ELAN 4 electro drill tool (from software version 
3.00)

GA849

ELAN 4 electro craniotome and multifunctional 
handpiece (2-ring)

GA860

ELAN 4 electro MIS hand piece

GA861

ELAN 4 electro Standard handpiece (1-ring) L4

GA862

ELAN 4 electro Standard handpiece (1-ring) L7

GA863

ELAN 4 electro Standard handpiece (1-ring) L10

GA864

ELAN 4 electro Standard handpiece (1-ring) L13

GA865

ELAN 4 electro Standard handpiece (2-ring) L7

GA866

ELAN 4 electro Standard handpiece (2-ring) L10

GA867

ELAN 4 electro Standard handpiece (2-ring) L13

GA868

ELAN 4 electro Standard handpiece (2-ring) L17

GA869

ELAN 4 electro Standard handpiece (2-ring) L22

SOP-AIC-
5001568 
(TA014401)

Instructions for use of ELAN 4 electro control unit 
GA800

Document No.:  IFU   -  Version: 3.0  -  Document ID: SOP-AIC-5001568   Date/Time Printed/Viewed:  2022-04-13  15:59 (CET)

Effective

Summary of Contents for ELAN 4

Page 1: ...ESCULAP USA Instructions for use Technical description ELAN 4 electro control unit GA800 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Page 2: ...11 12 13 14 11 15 16 18 17 1 2 3 4 5 6 7 8 9 10 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Page 3: ...19 20 21 22 23 24 25 27 28 29 30 26 29 28 30 31 GA810 GA808 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Page 4: ...12804 GA395SU GA808 GA849 GA822 GA861 GA864 GA865 GA869 GA860 GA836 GA831 GA837 GA832 GA824 GA833 GA844 GA810 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 1...

Page 5: ...tton 29 Pedal 30 Rotational direction button for motor 31 Speed range limit button only for wireless foot control GA810 Symbols on product and packaging Caution Observe important safety information su...

Page 6: ...g L10 GA867 ELAN 4 electro Standard handpiece 2 ring L13 GA868 ELAN 4 electro Standard handpiece 2 ring L17 GA869 ELAN 4 electro Standard handpiece 2 ring L22 Drill GA844 ELAN 4 electro power drill Sy...

Page 7: ...eed up Slow down Oscillation mode Oscillation angle Flow rate Increase the value Reduce the value System volume Display brightness System language Reset device to factory settings Go back to menu stru...

Page 8: ...control unit 11 6 2 Function checks 11 6 3 Safe operation 11 6 3 1 Activating the motor pump control panel 12 6 3 2 Changing the maximum speed setting rotational direction of the applied part 12 6 3 3...

Page 9: ...al circulatory system Note The safe and effective use of electrically driven applied parts depends heav ily on influences that only the user can control him or herself Therefore the specifications pro...

Page 10: ...ferent applied parts and has one connection socket for a foot control Only one applied part can be operated at a time Note The essential performance features are defined by the speed and rota tional d...

Page 11: ...t the device via the potential equalization terminal at the rear panel of the device to the potential equalization system of the room used for medical purposes Note The potential equalization lead can...

Page 12: ...n of the product from the main power supply is only guaranteed when the power cord is unplugged Switching off the product Activate the power OFF switch 8 Unplug the power cord from the power socket 13...

Page 13: ...mp tube does not get clamped in Insert the bottle holder 16 for the sterile fluid bottle in bottle holder mount 15 Insert the punching spike into the sterile fluid bottle When using sterile fluid bott...

Page 14: ...o be used are in good working order Check the secure connection of all products to be used Ensure that the parameters are set and the unit is operated according to the instructions for use and the saf...

Page 15: ...3 3 Activating the coolant and irrigation pump Flush function continuous flushing Activate the pump control panel see Chapter 6 3 1 Activate the coolant and irrigation pump Activate the button for sw...

Page 16: ...wise rotation on the control unit The motor direction indicator for counterclockwise rotation is illumi nated in the applied part control panel Activate the control element of the hand control The app...

Page 17: ...les c Press button import or export Follow the instructions on the display Note Profiles must be saved after creation or modification Save Profile Press button 6 4 2 Motor settings The motor settings...

Page 18: ...l number of the wireless foot control and select Yes to con firm the message Disconnect the GA810 wireless foot control by pressing the Disconnecting from GA810 button or switch off the control unit f...

Page 19: ...nfecting agents Only use cleaning disinfecting agents approved for surface clean ing Follow the manufacturer s instructions for the respective cleaning disinfecting agent Do not clean the product in a...

Page 20: ...rol defective Replace foot control Critical error in manual control module Unexpected error occurred in the area of the hand control If this occurs repeatedly please replace the hand control Unexpecte...

Page 21: ...test Release the activation If the pedal or button is not activated the foot control is defective If necessary replace foot control Multiple control elements of the foot control acti vated Multiple co...

Page 22: ...applied part Coupling processes cannot be per formed Release the release button on the motor cable between connecting dis connecting and releasing locking the applied part and activate again Motor cou...

Page 23: ...art operated in counter clockwise rotation Serrated tool run in counterclock wise rotation mode Operate serrated tool in clockwise rotation No central alignment of the crani otomy reamer to the dura p...

Page 24: ...ap Repair Hotline Phone 1 800 214 3392 Fax 1 314 895 4420 Other service addresses can be obtained from the address indicated above 11 Accessories Spare parts 11 1 ELAN 4 electro motor cables applied p...

Page 25: ...piece IIa GA861 ELAN 4 electro Standard handpiece 1 ring L4 IIa GA862 ELAN 4 electro Standard handpiece 1 ring L7 IIa GA863 ELAN 4 electro Standard handpiece 1 ring L10 IIa GA864 ELAN 4 electro Stand...

Page 26: ...Operation Storage and transport Temperature 10 C to 40 C 10 C to 50 C Relative humidity 30 to 75 10 to 90 Atmospheric pressure 700 hPa to 1 060 hPa 500 hPa to 1 060 hPa The recycling pass can be downl...

Page 27: ...n Germany Phone 49 0 7461 95 0 Fax 49 0 7461 95 26 00 www bbraun com AESCULAP a B Braun brand SOP AIC 5001568 TA014401 2021 07 Rev 3 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time P...

Reviews: