Aesculap ELAN 4 Instructions For Use/Technical Description Download Page 13

10

USA

6.1.8

Connecting the ELAN 4 electro motor cable 
GA804/GA805/GA806 or ELAN 4 electro drill tool GA844 to 
control unit

Note

The motor cable is sterile. Sterile separation occurs at the motor cable.

Note

The plug connection of the motor cable has a blue coded ring and a hollow 
dot.

Position connector for control unit 

c

 on the motor cable so that mark-

ing 

b

 on the connector is aligned with marking 

a

 on the connection 

socket for applied parts 

7

, see Fig. 3.

Plug connector for control unit 

c

 on the motor cable firmly into one of 

the two connection sockets for applied parts 

7

.

Fig. 3

Connecting the motor cable

Legend 

a

Marking for connection socket

b

Marking for connector

c

Connector for control unit

Note

The motor cable must be connected to the control unit without an applied 
part or with the applied part locked ("Off" position).

Otherwise the control unit will not recognize the applied part, and a mes-
sage will appear on the display (except for drill tool GA844).

If the applied part is not recognized;

– Lock the applied part, see Chapter 6.1.12.

– Release the applied part again, see Chapter 6.1.11.

6.1.9

Connecting the ELAN 4 electro single-use tube set GA395SU

Note

The tube set is sterile. Sterile separation occurs at the tube set.

Open flap 

a

 of the coolant and irrigation pump 

3

, see Fig. 4.

Insert single-use tube set 

c

:

– Place pump tube with loop over the roller wheel 

b

.

– Push nib 

e

 of the single-use tube set under the pull slide 

d

 until the 

nib engages.

Close flap of the coolant and irrigation pump 

a

. When doing this, make 

certain the pump tube does not get clamped in.

Insert the bottle holder 

16

 for the sterile fluid bottle in bottle holder 

mount 

15

.

Insert the punching spike into the sterile fluid bottle.

When using sterile fluid bottles made from glass: Open the ventilation 
flap on the punching spike.

Hang the sterile fluid bottle from the bottle holder 

16

.

Fasten the tube to the motor cable using fixation clamps.

Shorten the length of the tubes to correspond with the applied part 
being used and connect to the spray nozzle.

Fig. 4

Connecting the single-use tube set 

Legend 

a

Flap of the coolant and irrigation pump

b

Roller wheel

c

Single-use tube set

d

Pull slide

e

Nib of the single-use tube set

6.1.10 Connecting the applied part to the motor cable

Note

All ELAN 4 electro motor cables (GA804/GA805/GA806) can be used in 
combination with all applied parts that do not have a fixed connection with 
a cable.

Connect connector for motor cable 

24

 to connector for application 

part 

19

 of the motor cable. Ensure that the nose 

23

 on the motor cable 

is aligned with the groove on the coupling of the application part.

The applied part engages. When the visual field is “Off” 

20

 on the 

motor cable, a gold-colored marking is visible.

The control unit 

1

 recognizes the applied part type and displays this 

type in the corresponding applied part control panel in the display 

2

.

The most recent settings for this applied part type at the same motor 
connection socket are shown on the display 

2

.

Note

The applied part connected to this motor cable is only ready for operation if 
a gold-colored marking is visible in the visual field "On" 

21

 on the motor 

cable.

a

b

c

a

b

c

d

e

Document No.:  IFU   -  Version: 3.0  -  Document ID: SOP-AIC-5001568   Date/Time Printed/Viewed:  2022-04-13  15:59 (CET)

Effective

Summary of Contents for ELAN 4

Page 1: ...ESCULAP USA Instructions for use Technical description ELAN 4 electro control unit GA800 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Page 2: ...11 12 13 14 11 15 16 18 17 1 2 3 4 5 6 7 8 9 10 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Page 3: ...19 20 21 22 23 24 25 27 28 29 30 26 29 28 30 31 GA810 GA808 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Page 4: ...12804 GA395SU GA808 GA849 GA822 GA861 GA864 GA865 GA869 GA860 GA836 GA831 GA837 GA832 GA824 GA833 GA844 GA810 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 1...

Page 5: ...tton 29 Pedal 30 Rotational direction button for motor 31 Speed range limit button only for wireless foot control GA810 Symbols on product and packaging Caution Observe important safety information su...

Page 6: ...g L10 GA867 ELAN 4 electro Standard handpiece 2 ring L13 GA868 ELAN 4 electro Standard handpiece 2 ring L17 GA869 ELAN 4 electro Standard handpiece 2 ring L22 Drill GA844 ELAN 4 electro power drill Sy...

Page 7: ...eed up Slow down Oscillation mode Oscillation angle Flow rate Increase the value Reduce the value System volume Display brightness System language Reset device to factory settings Go back to menu stru...

Page 8: ...control unit 11 6 2 Function checks 11 6 3 Safe operation 11 6 3 1 Activating the motor pump control panel 12 6 3 2 Changing the maximum speed setting rotational direction of the applied part 12 6 3 3...

Page 9: ...al circulatory system Note The safe and effective use of electrically driven applied parts depends heav ily on influences that only the user can control him or herself Therefore the specifications pro...

Page 10: ...ferent applied parts and has one connection socket for a foot control Only one applied part can be operated at a time Note The essential performance features are defined by the speed and rota tional d...

Page 11: ...t the device via the potential equalization terminal at the rear panel of the device to the potential equalization system of the room used for medical purposes Note The potential equalization lead can...

Page 12: ...n of the product from the main power supply is only guaranteed when the power cord is unplugged Switching off the product Activate the power OFF switch 8 Unplug the power cord from the power socket 13...

Page 13: ...mp tube does not get clamped in Insert the bottle holder 16 for the sterile fluid bottle in bottle holder mount 15 Insert the punching spike into the sterile fluid bottle When using sterile fluid bott...

Page 14: ...o be used are in good working order Check the secure connection of all products to be used Ensure that the parameters are set and the unit is operated according to the instructions for use and the saf...

Page 15: ...3 3 Activating the coolant and irrigation pump Flush function continuous flushing Activate the pump control panel see Chapter 6 3 1 Activate the coolant and irrigation pump Activate the button for sw...

Page 16: ...wise rotation on the control unit The motor direction indicator for counterclockwise rotation is illumi nated in the applied part control panel Activate the control element of the hand control The app...

Page 17: ...les c Press button import or export Follow the instructions on the display Note Profiles must be saved after creation or modification Save Profile Press button 6 4 2 Motor settings The motor settings...

Page 18: ...l number of the wireless foot control and select Yes to con firm the message Disconnect the GA810 wireless foot control by pressing the Disconnecting from GA810 button or switch off the control unit f...

Page 19: ...nfecting agents Only use cleaning disinfecting agents approved for surface clean ing Follow the manufacturer s instructions for the respective cleaning disinfecting agent Do not clean the product in a...

Page 20: ...rol defective Replace foot control Critical error in manual control module Unexpected error occurred in the area of the hand control If this occurs repeatedly please replace the hand control Unexpecte...

Page 21: ...test Release the activation If the pedal or button is not activated the foot control is defective If necessary replace foot control Multiple control elements of the foot control acti vated Multiple co...

Page 22: ...applied part Coupling processes cannot be per formed Release the release button on the motor cable between connecting dis connecting and releasing locking the applied part and activate again Motor cou...

Page 23: ...art operated in counter clockwise rotation Serrated tool run in counterclock wise rotation mode Operate serrated tool in clockwise rotation No central alignment of the crani otomy reamer to the dura p...

Page 24: ...ap Repair Hotline Phone 1 800 214 3392 Fax 1 314 895 4420 Other service addresses can be obtained from the address indicated above 11 Accessories Spare parts 11 1 ELAN 4 electro motor cables applied p...

Page 25: ...piece IIa GA861 ELAN 4 electro Standard handpiece 1 ring L4 IIa GA862 ELAN 4 electro Standard handpiece 1 ring L7 IIa GA863 ELAN 4 electro Standard handpiece 1 ring L10 IIa GA864 ELAN 4 electro Stand...

Page 26: ...Operation Storage and transport Temperature 10 C to 40 C 10 C to 50 C Relative humidity 30 to 75 10 to 90 Atmospheric pressure 700 hPa to 1 060 hPa 500 hPa to 1 060 hPa The recycling pass can be downl...

Page 27: ...n Germany Phone 49 0 7461 95 0 Fax 49 0 7461 95 26 00 www bbraun com AESCULAP a B Braun brand SOP AIC 5001568 TA014401 2021 07 Rev 3 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time P...