Guidance and manufacturer’s declaration –
electromagnetic immunity
The Aeroneb Pro nebulizer system is intended for use in the
electromagnetic environment specified below. The customer or the
user of the Aeroneb Pro nebulizer system should assure that it is used
in such an environment.
Immunity
test
IEC/EN
60601
Test level
Compliance
level
Electromagnetic
environment
–
guidance
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 Vrms
80 MHz to
2.5 GHz
[3]V
[3]V
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Aeroneb
Pro, including cables,
than the
recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
Recommended
separation distance
d = [1.17] √P
d = [1.17] √P...
80 MHz to 800 MHz
d = [2.33] √P...
800 MHz to 2.5 GHz