Aeroneb
®
Professional Nebulizer System Instruction Manual
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
EN 61000-4-11
< 5% Ut
(> 95% dip
in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
< 5% Ut
(> 95% dip
in Ut)
for 5 sec
< 5% Ut
(> 95% dip
in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
< 5% Ut
(> 95% dip
in Ut)
for 5 sec
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
If the user of
the Aeroneb Pro
requires continued
operation during
power mains
interruption, it is
recommended
that the Aeroneb
Pro be powered
from an
uninterruptible
power supply
or battery.
Power
frequency
(50/60 Hz)
Magnetic field
EN 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at
levels
characteristic of
a typical location
in a typical
commercial or
hospital
environment.
Note: Ut is the a.c. mains voltage prior to application of the test level.