Aeroneb
®
Professional Nebulizer System Instruction Manual
Guidance and manufacturer’s declaration –
electromagnetic immunity
The Aeroneb Pro nebulizer system is intended for use in the
electromagnetic environment specified below. The customer or the
user of the Aeroneb Pro nebulizer system should assure that it is used
in such an environment.
Immunity test
IEC/EN
60601 Test
level
Compliance
level
Electromagnetic
environment
–
guidance
Electrostatic
discharge
(ESD)
EN 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be
wood, concrete or
ceramic tile.
If floors are
covered with
synthetic material,
the relative
humidity should
be at least 30%.
Electrical fast
Transient/burst
EN 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Surge
EN 61000-4-5
±1 kV
differential
mode
±2 kV
common
mode
±1 kV
differential
mode
±2 kV
common
mode
Mains power
quality should be
that of a typical
commercial or
hospital
environment.