5.
Start the measurement by pressing the
START/STOP
button.
W
= 2. .d.R
S-ES
To display the measurement parameters, press
DISPLAY
, R
, d, U
, I
and its frequency, U and its frequency).
To measure the resistances of the auxiliary electrodes
H (Z)
,
S (Y)
,
ES (Xv)
and
E
(X)
, or if the resistance of the electrodes is too high (see § 10.1), start the measure
ment by a long press of the
START/STOP
button.
R and R , then R
and R will be displayed.
> 2s
1.
Set the switch to
.
2.
Place electrodes
S (Y)
and
ES (Xv)
at a distance
A
apart, then place elec
trodes
H (Z)
and
E (X)
in the same straight line, at a distance
d
measured from
the midpoint of distance
A
. Distance
d
must be between 2 and 30m.
3.
p
of not more than
1/20 of d
.
4.
Connect the cables to the electrodes, then to terminals
H (Z)
,
S (Y)
,
ES (Xv)
,
and
E (X)
.
5.
Program the distance into the instrument as described in § 5.6.3.