Note:
Before you calibrate the A/D or D/A function, power must have been
turned on for at least
15 minutes
to make sure the DAS card is stable.
A calibration utility,
AutoCali.EXE
, is included on the companion
CD-ROM :
AutoCali.EXE
MIC-3716 calibration utility
This calibration utility is designed for the Microsoft Windows
environment. You can access this program from the default location:
\Program Files\ADVANTECH\ADSAPI\Utilities\MIC3716
5.2 VR Assignment
There is one variable resistor (VR1) on the MIC-3716 to adjust the
accurate reference voltage on the MIC-3716. We have provided a test
point (See TP4 in
Fig. 5-1
) for you to check the reference voltage on
board. Before you start to calibrate A/D and D/A channels, please
adjust VR1 until the reference voltage on TP4 has r5.0000 V.
Fig. 5-1
shows the locations of VR1 and TP4.
Summary of Contents for MIC-3716
Page 2: ...ii This page is left blank for hard printing...
Page 6: ...vi This page is left blank for hard printing...
Page 8: ...viii Table E 2 D A binary code table 117...
Page 11: ...1 Introduction 1 CHAPTER...
Page 17: ...7 Fig 1 1 Installation Flow Chart...
Page 21: ...11 Installation and Configuration CHAPTER 2...
Page 40: ...30 This page is left blank for hard printing...
Page 41: ...31 Signal Connections CHAPTER 3...
Page 44: ...34 Fig 3 1 I O connector pin assignments for the MIC 3716...
Page 50: ...40...
Page 54: ...44 This page is left blank for hard printing...
Page 55: ...45 Software Programming Overview CHAPTER 4...
Page 60: ...50 This page is left blank for hard printing...
Page 61: ...51 Calibration CHAPTER 5...
Page 77: ...67 Appendixes...
Page 81: ...71 Appendix B Block Diagrams...
Page 82: ...72 This page is left blank for hard printing...
Page 120: ...110 This page is left blank for hard printing...