Temperature Control Units : Sentra with VE Series Instrument
Page: 17
Advantage Engineering, Inc.
525 East Stop 18 Road Greenwood, Indiana 46142
317-887-0729 Fax: 317-881-1277
Service Department Fax: 317-885-8683
www.AdvantageEngineering.com
Use of incorrect voltage will void the unit’s warranty and cause a significant
hazard that may result in serious personal injury and/or unit damage.
2.
For standard units with 10 and 16 KW heaters and up to 3 horsepower pumps, a
four conductor cable, 10 foot in length, is provided for connection to an operator
supplied fused disconnect.
3.
For units with 24 and 34 KW heaters, the operator must provide a four conductor
power cable and the fused disconnect.
4.
The owner supplied fused disconnect must be sized and installed according to
the unit’s power supply requirements and local electrical codes.
B.
Models with Factory included Disconnect Switch and other Custom Features
1.
Some units may be customized and include a factory supplied power disconnect
switch and/or higher specification electrical enclosure. Electrical power supply
requirements are identified on the equipment data tag. Verify that available
voltage supply is the same as the unit’s voltage requirements.
WARNING: DO NOT connect the unit to a voltage supply source not equal
to the unit’s voltage requirements as specified on the unit’s data plate. Use
of incorrect voltage will void the unit’s warranty and cause a significant
hazard that may result in damage to the unit or serious personal injury.
2.
Appropriate conduit and fittings should be selected which will maintain the
integrity of the cabinet.
3.
Supply a power conductor sized according to the unit’s power supply
requirements. Connect the power conductor to
the unit’s power supply entry terminal block.
C.
Control Circuit Wiring
1.
The unit’s supplied control circuit is 110 volt, 1
phase, 60 cycle. The control circuit is supplied
by the factory installed transformer. A control
circuit fuse is provided.
WARNING: Do not connect the unit to a voltage supply not equal to the unit’s voltage
requirements as specified on the unit’s data plate. Use of incorrect voltage will void the unit’s
warranty and cause a significant hazard that may result in serious personal injury and unit
damage.
WARNING: Electric Shock Hazard. High Voltage is present in the electrical cabinet.
Disconnect power before servicing. Follow all facility lock-out tag-out procedures.
Typical control circuit transformer
fuse
Summary of Contents for Sentra SK Series
Page 2: ......
Page 6: ...Page 6 THIS PAGE INTENTIONALLY BLANK ...
Page 20: ...Page 20 THIS PAGE INTENTIONALLY BLANK ...
Page 30: ...Page 30 THIS PAGE INTENTIONALLY BLANK ...
Page 36: ...Page 36 THIS PAGE INTENTIONALLY BLANK ...
Page 68: ...Page 68 THIS PAGE INTENTIONALLY BLANK ...
Page 69: ...END 2017 ADVANTAGE ENGINEERING INC RE 01272017 ...
Page 70: ......