Table F‑2. Converter box and analog box UPP commands (Continued)
Parameter
Command
Description
Sensor head for
analog output
<AA>ax<VAL>
For analog box with
multiple connections,
specify the analog out:
<AA>axsX<VAL>
Sets the sensor head that will be used for analog
output of the converter box.
Valid values are N1 to N8 (sensor head number)
or A1 to A8 (sensor head address).
Address
<AA>ga<VAL>
Sets the address of the converter box. The
pyrometer will automatically reset and begin
working with the new address.
Valid values are 00 to 99.
• 00 to 97 = Valid addresses
• 98 = Global address. Use this address
to send entry commands to all the
connected pyrometers. The pyrometers will
not respond to confirm that a change has
been made.
Example:
98fh1
sets the temperature
display of all the connected pyrometers to
°F.
• 99 = Override address. When only one
pyrometer is connected and the converter
box address is unknown, use this address
to send entry commands to the connected
pyrometer. The pyrometer will respond to
confirm that a change has been made.
Baud rate
<AA>br<VAL>
Sets the baud rate for the converter box (RS-232
and RS-485 interface).
Valid values are 2 to 6 and 8.
2 = 4800 baud
3 = 9600 baud
4 = 19200 baud
5 = 38400 baud
6 = 57600 baud
7 = Not allowed
8 = 115200 baud