background image

©2021 Advanced Energy Industries, Inc.

Advanced Energy’s Artesyn iLS600, iLS600-R and iLS1500 

series are programmable DC power supplies with a single 

output that offers output power to 600 and 1500 watts. With 

12-bit D/A & A/D converters embedded, the power supplies 

come with the capability of reporting voltage and current very 

accurately. The iLS600, iLS600-R and iLS1500 series provide 

convenient digital rotary controls for voltage and current 

adjustment and the power supplies also come with rear ports 

that allow remote control via USB, Ethernet, and analog 

control inputs. The iLS600, iLS600-R and iLS1500 series are 

LXI certified.

ARTESYN

iLS600, iLS600-R and iLS1500 Series

Intelligent Laboratory Power
600W/1500W Programming Manual

AT A GLANCE 

Total Power

600 and 1500 Watts

# of Outputs

Single

USER MANUAL

Summary of Contents for iLS1500 Series

Page 1: ...ltage and current very accurately The iLS600 iLS600 R and iLS1500 series provide convenient digital rotary controls for voltage and current adjustment and the power supplies also come with rear ports...

Page 2: ...nel 6 1 0 10 Mode Transitions Commanded via Remote SCPI Command 7 1 1 Local Mode 8 1 2 Remote Mode 9 1 2 1 SCPI Commands 9 1 2 2 Front Panel Commands 9 1 2 3 Additional Capabilities 10 1 3 Remote with...

Page 3: ...nal SCPI Commands 15 1 6 6 Additional Capabilities 15 2 REMOTE SENSE CONFIGURATION 16 2 0 Enable Disable BENCH iLS Remote Sense 16 2 1 Calculate the Lead Resistance 16 3 CALIBRATION 17 3 0 Calibration...

Page 4: ...re Status Register 24 6 0 11 Hardware Status Register 25 6 0 12 Operation Status Register 26 6 0 13 Standard Event Status Register 28 6 0 14 Error Event Queue 29 6 0 15 Error Condition Register 31 6 0...

Page 5: ...ility is disabled That is the Enable button cannot be used either to enter Setup or to deactivate the output As in Remote mode the Analog Voltage and Analog Current inputs on the Interface board are i...

Page 6: ...activated and this output is proportional to the output voltage of the power supply When this output voltage is fed to one or more analog inputs of other supplies and the analog inputs of those suppl...

Page 7: ...em MODe LOCal For Local mode SYSTem MODe REMote For Remote mode SYSTem MODe RWLock For Remote with Lock mode SYSTem MODe VOLTage For Analog Voltage Control mode SYSTem MODe CURRent For Analog Current...

Page 8: ...ction level amps CURRent PROTect Query current protection level amps POWer PROTect value Set power protection level watts POWer PROTect Query power protection level watts OUTPut AUTO ON OFF Enable dis...

Page 9: ...voltage volts MEASure CURRent Measure and return output current amps VOLTage PROTect value Set voltage protection level volts VOLTage PROTect Query voltage protection level volts CURRent PROTect valu...

Page 10: ...ck Mode which cannot be exited from the Front Panel will not be stored in non volatile memory An attempt to do so will replace it with Remote Mode which can be exited from the Front Panel 1 4 Analog V...

Page 11: ...nt setpoints using the commands MEASure VOLTage Measure and return output voltage volts MEASure CURRent Measure and return output current amps OUTPut Query output state VOLTage Query voltage setpoint...

Page 12: ...Mode is entered via the front panel you will be asked to select which scaling you want You may also select scaling with the SCPI command SYSTem MODe ASCale CURRent 3 5 10 1 5 2 Voltage Setpoint The Vo...

Page 13: ...the output voltage and current 1 6 Analog Dual Control Mode 1 6 1 Voltage Setpoint When the unit is in Analog Dual Control Mode the Analog Voltage channel supplies the Voltage Setpoint Analog Input Sc...

Page 14: ...via the front panel you will be asked to select which scaling you want You may also select scaling with the SCPI command SYSTem MODe ASCale CURR 3 5 10 1 6 3 Activating deactivating the output You ca...

Page 15: ...y Status and Error registers as described in Section 6 0 1 6 6 Additional Capabilities When the output is OFF you can switch operating modes as described in Section 1 0 When the output is ON you will...

Page 16: ...enable Remote Sense RSENse ON 3 Execute the command to calculate lead resistance RSENse RESistance CALCulate current Initiates calculation The current value is the current in amperes at which the lea...

Page 17: ...ltage to a high value recommend 90 of the rated output voltage Record the high voltage setpoint 4 Use the DMM to record the actual output voltage of the supply Record the high voltage measured 5 Comma...

Page 18: ...rgy com 18 10 Command BENCH iLS to perform the current calibration based on the recorded values CALibration CALCulate CURRent low current setpoint low current measured high current setpoint high curre...

Page 19: ...power up sequence In addition you may exercise a number of Self Test options with the following SCPI commands TST Execute Self Test and return the results TEST SELFtest EXECute Same as above TEST QUER...

Page 20: ...interface boards and 4 the software revision number for the main and interface board processors For example a 100V 10A model may in response to IDN return Artesyn Power Bench 100 10 iLS 031418032001...

Page 21: ...eries Rev 05 20 21_ 1 0 advancedenergy com 21 6 0 Architecture Figure 1 shows the Overall Status Architecture Each section will then be discussed in greater detail SECTION 6 STATUS AND ERRORS Figure 1...

Page 22: ...ition inputs are 6 0 1 Over Voltage Error When activated the unit has exceeded the Over Voltage Protection limit 6 0 2 Over Current Error When activated the unit has exceeded the Over Current Protecti...

Page 23: ...utput Error Regulation of the output has been lost 6 0 8 SCPI Access Access to the Questionable Status Register is via the following commands STATus QUEStionable EVENt Rtn Event register STATus QUESti...

Page 24: ...emperature The primary board temperature surpasses its allowed operating limit Fan Stall The cooling fan is non operative SCPI Access Access to the Temperature Status Register is via the following com...

Page 25: ...ter four individual condition inputs are 12V Bias Error The internal 12V bias is either too low or too high for valid operation 3 3V Bias Error The internal 3 3V bias is either too low or too high for...

Page 26: ...QUEStionable HARDware ENABle value Set Enable register STATus QUEStionable HARDware ENABle Rtn Enable register 6 0 12 Operation Status Register Figure 5 shows the Operation Status Register The Questio...

Page 27: ...rrent mode where current is held constant at a specified setpoint and voltage will vary dependent on load Constant Power The output is in its constant power mode where power is held constant at a spec...

Page 28: ...individual condition inputs are Operation Complete This bit is not used by BENCH iLS Request Control This bit is not used by BENCH iLS Query Error This bit is not used by BENCH iLS Device Specific Err...

Page 29: ...s ESR Rtn Event register ESE value Set Enable register ESE Rtn Enable register 6 0 14 Error Event Queue The Error Event Queue buffers the occurrence of Errors in a FIFO queue The queue will store up t...

Page 30: ...r power 111 Output board over temperature 112 Primary board temperature error 113 Transformer temperature error 114 Fan stall error 121 PWM activation failure 122 Output error 131 12V bias error 132 3...

Page 31: ...itions Error Conditions The Error Conditions recognized by BENCH iLS identified by their error code are listed below If multiple error conditions are present the register will contain the sum of the i...

Page 32: ...uest Enable Register Figure 7 shows the Status Byte Register and the Service Request Enable Register The six individual condition inputs of the Status Byte Register are Error Event Queue When the Erro...

Page 33: ...t Summary Bit RQS Any of the four functioning bits of the Status Byte Register being set and enabled in the Service Request Enable Register will cause this bit to be set Operation Status If an Operati...

Page 34: ...B drivers are available on the Artesyn Power website at https artesynpower com drivers Be sure to download and extract the Bench USB Driver to your local drive The extraction location will be required...

Page 35: ...iLS600 iLS600 R and iLS1500 Series Rev 05 20 21_ 1 0 advancedenergy com 35 2 On the Driver tab select Update Driver 3 Select Browse my computer for driver software SECTION 7 DRIVERS...

Page 36: ...d iLS1500 Series Rev 05 20 21_ 1 0 advancedenergy com 36 4 Select Browse On the Browse for screen locate where the Bench USB Driver was extracted to and click next 5 On Windows Security select Install...

Page 37: ...0 21_ 1 0 advancedenergy com 37 6 Once the driver has been successfully installed windows will verify the installation and the Bench can now be accessed through USB communication 7 Device manager will...

Page 38: ...components Use the descriptions on the IVI foundation page referenced above to choose the components necessary for your installation Note that both the IVI and the IVI net components must both be inst...

Page 39: ...ers Prior to installing the LabView drivers LabView itself must be installed on your machine Once LabView is installed to install the Artesyn Power LabView Drivers download them from https artesynpowe...

Page 40: ...nt register of the Hardware Status Register a part of the Questionable Status Register structure The Error Event Queue ESE value This command sets the Standard Event Status Enable Register to the spec...

Page 41: ...scription of ESE If multiple bits are set the result is the sum of the set bit values ESR Returns the present contents of the Standard Event Status Event Register IDN Returns the identification string...

Page 42: ...alue The value is specified in decimal form with the following value assignments 1 Reserved not used 2 Reserved not used 4 Error Event Queue 8 Questionable Status 16 Message Available not used 32 Stan...

Page 43: ...2 measured2 This command uses the four parameters provided to compute slope offset calibration parameters for voltage control MEASure SCALar CURRent DC This command returns the most recently measured...

Page 44: ...ept for queries RSENse This command returns the present remote Remote Sense mode ON or OFF RSENse RESistance ohms Sets the remote sense resistance directly as opposed to running current through a lead...

Page 45: ...current protection threshold in amperes SOURce POWer LEVel IMMediate AMPLitude MIN MAX value This command sets the Power setpoint as specified The setpoint can be set to MIN 0 0 watts MAX model speci...

Page 46: ...Tectection LEVel This command returns the present over voltage protection threshold in volts STATus OPERation CONDition This command returns the present contents of the Operation Status Condition Regi...

Page 47: ...ts The returned value is in decimal form described in the description of STATus OPERation CONDition If multiple bits are set the result is the sum of the set bit values STATus PRESet Set the Enable Re...

Page 48: ...is the sum of the set bit values STATus QUEStionable EVENt This command returns the Questionable Status Event Register contents The returned value is in decimal form described in the description of ST...

Page 49: ...e result is the sum of the set bit values 1 Output Board Over Temperature Error 2 Primary Board Over Temperature Error 4 Fan stall Error STATus QUEStionable TEMPerature ENABle value This command sets...

Page 50: ...ng DCPSUPPLY WITH MEASURE is returned SYSTem CONFiguration SAVE This command saves configuration parameters to non volatile memory Contents consist of Analog current scaling select 3V 5V 10V Analog cu...

Page 51: ...r the SCPI command interface with no front panel capability VOLTage Analog Voltage Control The voltage setpoint is determined by the Analog Voltage input SCPI commands provide the rest of the control...

Page 52: ...4 Over Power Error 8 Output Board Over Temperature Error 16 Fan Stall Error 32 Output loss of regulation Error 64 12V Bias Error 128 3 3V Bias Error 256 Primary input Board Temperature Error 512 PFC...

Page 53: ...BENCH iLS the string 1999 0 is returned TEST SELFtest CLE This command clears Self Test results TEST SELFtest EXECute or TST Execute Self Test These commands execute Self Test and return the Self Tes...

Page 54: ...dustries including semiconductor equipment industrial manufacturing telecommunications data center computing and medical With deep applications know how and responsive service and support across the g...