iLS600, iLS600-R and iLS1500 Series
Rev. 05.20.21_#1.0
advancedenergy.com
11
1.4.2 Current Setpoint
The Current Setpoint is the last current value set when in Local Mode, and once in Analog
Voltage Control Mode, cannot be changed either from the front panel or by a SCPI
command.
1.4.3 Activating/Deactivating the Output
You can activate or deactivate the output either from the front panel Enable switch or by the
SCPI command:
OUTPut <ON|OFF>
Activate/deactivate output
1.4.4 Remote Queries
When the output is ON, you may query the output current and voltage, as well as the output
state and voltage and current setpoints using the commands:
MEASure:VOLTage?
Measure and return output voltage (volts)
MEASure:CURRent?
Measure and return output current (amps)
OUTPut?
Query output state
VOLTage?
Query voltage setpoint (volts)
CURRent?
Query current setpoint (amps)
POWer?
Query power setpoint (watts)
Voltage is returned in volts, current in amperes, and power in watts. The output state is
either ON or OFF.
1.4.5 Additional SCPI Commands
Regardless of the output state, you can query Status and Error registers, as described in
Section 6.0.
1.4.6 Additional Capabilities
When the output is OFF, you can switch operating modes, as described in Section 1.0.
When the output is ON, you will be able to execute only those commands which activate
and deactivate the output and measure the output voltage and current.
SECTION 1
OPERATING MODES