6.0.13 Standard Event Status Register
Figure 6 shows the Standard Event Status Register.
The Standard Event Status Register takes the standard SCPI structure. The eight individual
condition inputs are:
Operation Complete
This bit is not used by BENCH iLS.
Request Control
This bit is not used by BENCH iLS.
Query Error
This bit is not used by BENCH iLS.
Device Specific Error
This bit is set if a Device Specific Error occurs. These include all Questionable Register
events and Error Queue Overflow.
Execution Error
This bit is not used by BENCH iLS.
SECTION 6
STATUS AND ERRORS
Figure 6: Standard Event Status Register