iLS600, iLS600-R and iLS1500 Series
Rev. 05.20.21_#1.0
advancedenergy.com
40
SCPI Commands
*CLS
— This command clears the following registers:
•
The Standard Event Status Register
•
The Event register of the Questionable Status Register structure.
•
The Event register of the Operation Status Register structure.
•
The Event register of the Temperature Status Register (a part of the Questionable
Status Register structure).
•
The Event register of the Hardware Status Register (a part of the Questionable
Status Register structure).
•
The Error/Event Queue.
*ESE <value>
— This command sets the Standard Event Status Enable Register to the
specified value. The value is specified in decimal form with the following value assignments:
1
Operation Complete
2
Request Control (not used)
4
Query Error (not used)
8
Device-Specific Error – Set for:
•
Calibration Error
•
Error Queue Overflow
•
Hardware Error
12V Bias Error
3.3V Bias Error
PFC Failure Pending
PFC Failure
•
Output Error
•
Over-Voltage Condition
•
Over-Current Condition
•
Over-Power Condition
•
Self-Test Error
APPENDIX