iLS600, iLS600-R and iLS1500 Series
Rev. 05.20.21_#1.0
advancedenergy.com
8
1.1 Local Mode
In Local Mode, you will control BENCH iLS primarily using the front panel controls. However,
many SCPI commands are still available – only setting of setpoints and activating the output
are disallowed.
When the output is OFF, the set of available commands is:
VOLTage:PROTect <value>
Set voltage protection level (volts)
VOLTage:PROTect?
Query voltage protection level (volts)
CURRent:PROTect <value>
Set current protection level (amps)
CURRent:PROTect?
Query current protection level (amps)
POWer:PROTect <value>
Set power protection level (watts)
POWer:PROTect?
Query power protection level (watts)
OUTPut:AUTO <ON|OFF>
Enable/disable the Auto-Start feature
OUTPut:AUTO?
Query present Auto-Start state
SYSTem:CONFiguration:SAVE
Save configuration data to non-volatile memory
When the output is ON, the set of available commands is:
OUTPut?
Query output state
OUTPut OFF
Deactivate output
MEASure:VOLTage?
Measure and return output voltage (volts)
MEASure:CURRent?
Measure and return output current (amps)
Regardless of the output state, you can query Status and Error registers, as described in
Section 6.0.
SECTION 1
OPERATING MODES