Pump temperature for
start-up condition
Operation in local
mode
Operation in remote
mode
C 10
Pump operation
The ACP Series uses a synthetic
oil fluid.
The amount required for pump
operation is set at the factory. Do not modify this oil level.
The oil drain will be done during pump overhaul by the manufacturer's
trained personnel.
The pump is equipped with thermal sensors.
When switching on the pump, if the temperature is:
- less than 12°C,
- or over 40°C,
the pump doesn't start, but the fan is energized. The pump will start
automatically when the ambient temperature is back in the authorized
temperature range.
Auto-restart
When the pump is stopped with an overtemperature issue, it will
restart automatically when ambient temperature has decreased until
the restart value.
It is the responsibility of the user to take all the measures required
to prevent risks resulting from this type of operation. The user must
provide a device (integrated in the equipment/host tool) to warn or
to avoid this restart.
In local mode, the pump can run only if the cover
plug (delivered with
the pump) is fitted on the remote control connector).
Wiring of the Remote cover plug
DB 15 pins, male connector (soldered side view). Factory wired with appropriate
jumper for local operation.
Note
It's the origin of
Start/Stop
control which has priority on the speed
selection:
- when the Start/Stop is made by the remote control cover plug, the
selected speed corresponds to the dry contact setting,
- when the Start/Stop is made by RS485 serial link, the speed
corresponds to the speed set point defined on serial link.