B 30
Electrical connections
Rear panel of the
pump
■
In accordance with recommandations of EN 61010-1+ A2, the
following warning symbol is on the variator inside the pump.
Warning: risk of electrical shock.
Voltage or current hazard sufficient to cause shock. Disconnect and
lockout power before servicing. Any intervention must be done by
trained personnel only.
■
Electrical frequency converter is in accordance with CE standards
and offers the following voltage range:
Circuit breaker
■
The motor is equipped with an electrical frequency converter
which allows automatically low or high voltage pump running,
according to range voltage 110 V to 230 V, 50/60 Hz.
■
The pump supplying cable is provided with the pump delivered.
The earthing of the pump (frequency converter, covers, body of
the pump) is realized by the cable connected with the network
customer. The network customer should have himself a connection
in the ground.
■
A 6 A circuit breaker is recommended for high voltage,
230 VAC + 10 %.
■
A 12 A circuit breaker is recommended for low voltage,
110 VAC + 10 %.
2/2
Alcatel Vacuum Technology France - Manuel de l'Utilisateur - ACP 15 / 28 / 40 Series 2