7351 Series Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-33
Measurement Disable func-
tion
:INHibit:FUNCtion
:<function>? *1
Reads a status in which the function setting is disabled.
Reply:1 or 0 (1: Invalid, 0: Valid)
Trigger
*TRG
Trigger command
Measurement
operation
:MEASure?
Sets the DCV measurement function and reads the measured
data.
:MEASure:<function>?
*1
Performs the measurement once and reads the measured
value.
:READ?
Initializes the trigger system and reads the measured data.
:FETCh?
Only reads the measured data.
Reply: Refer to “Output Data Format”.
Measurement
data memory
:TRACe:STATe
{ON,1|OFF,0}
Store setting
ON, 1: ON
OFF, 0: OFF
:TRACe:NUMBer
<int>,<int>
Readout range setting
<int>: 0 to 19999
:TRACe:DATA?
Data reading
Refer to Section 6.7.2,“Output Data Format.”
:TRACe:DATA:POIN
ts?
Number of measurement results reading
Reply: ddddd
:TRACe:DATA:NUM
Ber?
Data range reading
Reply: ddddd
:TRACe:CLEar
Initialization
Trigger
system
Start
:INITiate
Exits the IDLE state and moves to the trigger waiting state.
Continue
:INITiate:CONTinu-
ous {ON,1|OFF,0}
ON, 1: CONTINUOUS ON
OFF, 0: CONTINUOUS OFF
:INITiate:CONTinu-
ous?
Reply: 1 or 0
Trigger abort
:ABORt
Forcibly moves to the IDLE state.
Trigger source
selection
:TRIGger:SOURce
{IMMediate|MAN-
ual|EXTernal|BUS}
IMMediate
MANual
EXTernal (7351A, 7351E + Option 03)
BUS
:TRIGger:SOURce?
Reply: IMM, MAN, EXT, or BUS
Trigger delay
:TRIGger:DELay
{<real>|MAXi-
mum|MINimum}
<real>: 0 to 3600 (seconds)
(0)
:TRIGger:DELay?
{MAXimum
|MINimum}
Reply: +d.ddddE
dd
Sampling
count
(1 trigger)
:SAMPle:COUNt
{<int>|MAXi-
mum|MINimum}
<int>: 1 to 20000
(1)
:SAMPle:COUNt?
{MAXimum|MINi-
mum}
Reply: ddddd
*1: <Function>: Specifies a function, which is not enclosed by double quotes, for "measurement function".
Item
SCPI command
Description
Initial
value