7351 Series Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-30
System
Self-Test
*TST?
Performs the self- test and reads the result.
(It takes time to perform the self-test. Read the result after
the self-test is complete.)
Reply
0 : Pass
1 : Fail
*8
Calibration
CAL0
Sets the calibration mode to OFF.
(Writes the calibration factor when leaving the calibration
mode.)
CAL1
Sets the calibration mode to ON.
CAL?
Reply: CAL0 or CAL1
-
XOUT
Cancels (sets to OFF) the calibration mode.
(Does not write the calibration factor.)
-
PCn
Enters the STD value. (Displayed count value)
n:0 to
999999
XDTn
Enters the STD value. (Displayed value)
n: Displayed value on STD
-
CMNT
"str"
Sets the calibration information.
Enclose the character strings by using (") or (').
Up to 50 characters (one byte alphanumeric characters) can
be stored.
-
CMNT? Reply: CMNT"xxxx
xxx"
-
Setting parameter
*SAVn
n:0 to 3
Saves the setting parameters in area [n] of the non-volatile
memory.
-
SINI
Sets the values when shipping to all area from [0] to [3].
-
*RCLn
n:0 to 3
Loads the setting parameters from area [n] of the non-volatile
memory.
-
RINI
Loads the values when shipping as the setting parameters.
-
Error log reading
ERR?
Reads the error details.
Reply:
ddd,"xxxxxxxxxx"
• Up to 20 error queues can be stored. The error is output
according to the FIFO system.
• If more than 20 errors occur, the last error is replaced with
-350, “Queue overflow”. Any errors after this are not
saved.
• If no error occurs, +000, “No error” is returned.
-
*8: Self-test items differ.
Item
Command
Description
Initial
value
R6451
Up to 50 characters
Error character strings
(Up to 80 characters)
Error code