mean error with 8mmHg standard deviation).
Notes on Blood Pressure Data
Any blood pressure reading can be affected by the measurement site, the position of the patient, exercise, or the
patient’s physiologic condition. Environmental or operational factors which can affect the performance of the device
and/or its blood pressure reading are pacemakers and common arrhythmias such as atrial or ventricular premature
beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-clampsia, renal diseases,
patient motion, trembling, and shivering.
Specifications SP0
2
Sensor
Function
ChipOx
Masimo M-LNCS DCI
Saturation Normal (no motion)
± 2%
(70-100%A
rms
)
± 2%
(70-100%A
rms
)
Saturation Motion
N/A
± 3%
(70-100%A
rms
)
Low Perfusion
N/A
SpO
2
+/-2%
Pulse Heart rate
20-300 BPM: ±3 BPM
± 3% BPM
Pulse Heart rate w/motion
20-300 BPM: ±3 BPM
± 5% BPM
Wavelength & Output Power
RED: 660nm @ 3.5-4.5mW
IR: 905nm @ 3.5-4.5mW
RED: 660nm @ ≤ 15mW
IR: 905nm @ ≤ 15mW
For more detailed information, see the instructions for use included with each device.
Specifications Temperature Sensor
Temperature accuracy
30.1 to 42.9 °C (86.2 – 109.2 °F)
Quick Mode (Oral): ± 0.3°C
Std. Mode (Axillary/Rectal): ± 0.1°C
Direct Mode: ± 0.1°C
Response Time
3-10 sec (Oral Quick Mode)
6-10 (Oral Standard Mode)
8-12 sec.(Axillary Mode)
10-14 sec.(Rectal Mode)
60-120 sec.(Direct Mode all sites)
Resolution
0.1°C