ADView 2 User Manual | 40
93-9005-00
rev 4
September 1, 2016
9005USB
Mini USB to USB Cable
9005DON
USB Wi-Fi Dongle
9005P
Thermal Printer with Paper
9005PAPER
Thermal Printer Paper
9005SCAN
L
2D
Barcode Scanner
12. Status Messages & Alarms
If the ADView 2 has a problem taking a measurement, you may see a code on the display screen.
The
icon will blink with the measurement display result and code.
Below are common error codes. Take action as directed on the screen, or as suggested in the table below.
Status Messages
Code Displayed
Action User Should Take
Blood Pressure:
1-2, 4, 87 & 88
Check that the cuff is in the correct position.
Check that the cuff is properly tightened and that the tubing is properly connected.
Check that there is no excessive clothing between the arm and cuff.
Check that the cuff applied is the correct size.
The patient may have been moving too much.
Take another BP reading.