ADView 2 User Manual | 41
93-9005-00
rev 4
September 1, 2016
3
No action possible. Values are outside the reportable range
.
85 & 89
Check that the hose has no sharp bends and is not pinched.
Check that the patient is not lying on the cuff.
Check that the cuff is in the correct position.
Take another BP reading.
86
If cancelled reading was unintended, allow cuff to deflate & restart reading.
If cancelled reading was intended, no action needed.
90, 91, 97-107
Power off and repower on the ADView 2.
If issue persists, call SMI Customer Service. Service Required
Temperature:
505 & 506
Make sure patient has not had any warm beverages recently (oral), and then retry
measurement.
501-504, 507-514,
516-517
Remove and then replace the probe in the well.
Retry reading.
Power off and repower on the ADView 2.
Retry reading. If issue persists, call customer service.
520-524 & 526
Power off and repower on the ADView 2.
Retry reading.
If issue persists, replace the probe. Retry reading.
If issue persists, replace module or call Customer Service.
530
Use in normal room temperature environment suitable to the device.
SpO
2
:
121-124, 133-136,
150-156
Remove sensor from the patient and ensure proper positioning. Attempt new measurement
ensuring proper protocol as described in Section 7.
Unplug sensor and reconnect.
If issue persists, power off and repower on the ADView 2.
Retry reading. If issue persists, call Customer Service.
126-132
137-149 & 157-192
Power off and repower on the ADView 2.
If issue persists, replace module or call Customer Service
Electronic Medical Records (EMR):
601-610 & 612,614
Check EMR configuration, wireless strength, and proximity. Call your biomedical engineer or
IT support.
611
Ensure proper barcode and patient ID scanned.