ADView 2 User Manual | 34
9. Using ADView 2 EMR and Memory Functions
Section 4, “Good to Know Before You Begin,” provides some basic information about the ADView 2’s EMR and
memory functions. Here are more details about how these work.
Memory Mode
When you power up the unit, the memory identifier appears at the top of the display beside the Memory Icon
.
This is the number that will be associated with the next set of measurement results that are sent to the device
memory.
Press the Memory Button
on the front of the device to save results to memory. The Memory Icon, Check
Mark icon and all values that are displayed on the screen (at the time the button is pressed) will flash before the
results are saved. The following values can be saved to memory.
Parameter Name
Field Type
Patient ID
Alphanumeric Text
Time Stamp
Numeric, HH:MM:SS
Date Stamp
Numeric, MM.DD.YYYY or DD.MM.YYYY
Memory Location
Numeric
Systolic BP
Numeric
Diastolic BP
Numeric
BP Type
Text: AVG or SM
Pulse Rate
Numeric
Pulse Rate Source
Text: BP or SpO
2
Mean Arterial Pressure
Numeric
SpO
2
Numeric
Temperature
93-9005-00
rev 4
September 1, 2016
Numeric