©Adam Equipment Company 2008
11
5.8
SWITCH BETWEEN KILOGRAM TO POUND
•
The scale can be switched from kilograms to pounds when
enabled in the user settings. Press the
[Kg/lb /Set]
key for 2
seconds to change. The LED below the
lb
symbol will be
turned on.
•
Press the
[Kg/lb /Set]
key for 2 seconds to change back.
5.9
HOLD FUNCTION (ANIMAL WEIGHING ).
•
The scale will automatically hold a weight value that is greater
than the minimum weight value when the filter is set from
2 to
9
. See section 8.6 Filter. The display will remain unchanged
even when the weight is increased or returns to zero.
The higher the value the stronger the filter will be for animal
weighing applications.
•
To un-Hold the display to continue weighing or return to zero
you may press the
[Zero]
key. If the weight returns below the
minimum weight setting, the next time the scale is loaded it
will un-hold automatically.
•
To disable this function return the filter setting to
0
or
1
. This
can be found in parameter settings directly after User
parameters.
5.10 COUNT OPERATION
If the count function is enabled as explained in section 8.0, the following
operation is available.
•
In weighing mode when zero is displayed, press the
[On/Off]
key to enter count mode.
•
The display will show –
SanP-
and the
Count
indicator will be on.