©Adam Equipment Company 2008
9
ACCUMULATION
5.4
ACCUMULATION CONTROL SETTING
NB. If printing control is set to 99 Manual then the weight output will be
printed with accumulation.
•
Press
[Kg/lb /Set]
twice, quickly.
•
Display will show auto-accumulation state as “
Aut 0
”
.
The rightmost digit =
0
means manual accumulation.
The rightmost digit =
1
means auto accumulation when load is stable.
The rightmost digit=
2
means auto save displaying value when load is
stable. When unload to less than minimum weighing
value, the last stable value will be accumulated
automatically.
•
Use the
[x10 /Inc]
key to select and the
[Kg/lb /Set]
key to
enter your selection and return to normal weighing.
Note:
In automatic accumulation state, the
Auto
indicator will be on.
5.5
MANUAL MEMORY ACCUMULATION
If the accumulation function is enabled (refer to section 7.0), when
weighing is stable, press
[Acc/
]
, the accumulated value and the
number of accumulation will be displayed as “
n 26
”
. The value of “
n
” is
the number of weights that are stored. If communication address is 99,
the weighing value will be printed.
Note: Accumulation can only be used once in one weighing process.
The next accumulation is available only after displaying value is less
then minimum weighing value set in user parameters section 7.0.
Alternatively the indicator can be set to do an automatic memory
accumulation, see section below.