37
|
P a g e
© Adam Equipment Company 2017
8.0 PRINT
AND
STORE
RESULTS
If the test parameters are set to allow the test results to be printed (see
section 6.1.1) then the results of the test are printed both before and after the
test is complete. The initial part of the printout will show the test conditions. If
the test printing parameter has been set to complete then the results will be
printed at the interval time. When a test is completed, the user can print a
result from the test. At the same time as the printing is happening the results
are stored in the PMB memory. The results can be recalled from the memory
for printing over the serial interface.
8.1 PRINTING
FINAL
TEST
RESULTS
While the final values for the test are being displayed after the test has
ended, press the
[Print/
]
key to output the final results. The format of
these results are the same as the last lines of the complete printout:
LAST TEMP: 111C
TEST TIME: 05:30 Min
FINAL MASS: 11.820 g
MASS LOSS: 0.521 g
RESULT: 4.27 %M
8.2 STORING
TEST
RESULTS
If the test settings include the printing of results being enabled the analyzer
will store the results of the test into memory. (See section 6.1.1) The test
results will be stored using the file name given to it automatically by the PMB.
In the form RESxxxxx
The PMB is capable of holding up to 99 test results. If the analyzer memory
is full the display will show an MEM FULL message. Delete some of the
stored results to make room for new test results. See section 8.4.