49
|
P a g e
© Adam Equipment Company 2017
with the timer running. The timer will count up to 10:00 to allow time for the
chamber to come to a steady (room) temperature. During this time observe
the temperature in the chamber.
During this time the beeper will sound every minute to remind you that a
calibration is in progress.
Note the temperature inside the chamber. When the temperature is stable
and constant for some time press
[Setup/Enter]
to enter the lower
temperature value.
The display will change to allow the value to be entered.
Using the numeric entry method, enter the temperature inside the chamber,
for example 23ºC you would enter:
Press
[Setup/Enter]
to continue the test.
The heaters will turn on and begin heating the test chamber. The timer will
show the elapsed time as before. The chamber will be heated to
approximately 150ºC.
At the end of 30 minutes the beeper will sound, read the temperature in the
chamber. Press
[Setup/Enter]
to enter the second temperature value.
FIRST
00:00
FIRST
000
FIRST TEMP
023