20
|
P a g e
© Adam Equipment Company 2017
End Point Determination (manual stop active in all tests)
Time, Test ends at the end of a preset time.
Stable result, test ends if the results are stable within a value set by the
user.
Time or stable result, combines both methods. Stop whichever occurs
first, a stable result or the end of a preset time.
For timed test the user sets a time of typically 5 minutes to 99 minutes
Set stable results criteria. The test will stop when the weight changes less
than a preset amount in a preset time. The default setting is 0.002g in 15
seconds for PMB 53, and 0.02g in 15 seconds for PMB 202. The user can set
other values depending upon model.
Start Criteria
Start Manually when the
[Start]
key is pressed
Start Automatically when cover lowered.
Delays 5 seconds after cover is down before beginning.
Print settings:
To set if the analyzer should print and store results and when printing if the
compete test results or only the summary is printed and stored in the
analyzers internal memory.
Default settings (stored in PST 00, see section 6.3) are:
% Moisture
Single Temperature = 110ºC
Rapid = Off
Interval Time = 5 seconds
End Point Determination = Time and stable result
Time 60 minutes, and
Stable PMB 53 = 0.002g/15 seconds or Stable PMB 202 = 0.02g/15 seconds
Start Criteria = manual start.
Print Test = on
Print Format = Complete