On the inside of the scale, disconnect the
battery, display and power connectors from
the circuit board
Once disconnected, unscrew the board from the
base in the 3 designated locations
Once unscrewed, simply lift the board from the scale
and reverse these steps to replace with a new board.
Serial board
It is also possible to remove/ replace the serial board. After the main circuit board has been
removed from the scale, unscrew the singular screw fixing the serial board in place and lift
out the serial board.
8.0 Calibration/ Linearity settings
Follow these steps to calibrate the scale:
Turn on the power and wait for the scale to configure.
Before configuration is completed, hold down the
[Tare]
↵
key to bring up the user
information screen with username and password.
Entering the correct password will bring you to the setup menu.
3 PCB screw locations
Serial board screw
USB
Charging board