17
14.0
Troubleshooting
14.1
ALCOSIM simulator – troubleshooting checklist
In the event of a calibration failure, first check that:
The plastic tubing is not overused, and it is free of condensation
The alcohol reference solution value is 50 mg/dL (0.50 g/L)
The alcohol reference solution is not expired and has not been in the ALCOSIM simulator for more than 5 days
All connections to and from the ALCOSIM simulator are correct and secure
14.2
ALCOSIM simulator – troubleshooting table
If
Then
There is a leak in the seal.
The rubber ”O” ring seal is damaged or misaligned. Remove the top
housing and reposition the seal. Replace the seal if it is worn or
damaged. Call service if you need a replacement “O” ring.
The ALCOSIM simulator does not reach or
maintain a solution temperature of 34°C.
The thermostat or heating element may be defective. Send for service.
There is condensation in the plastic tubing.
Fit the plastic tubing on an aquarium air pump, or similar setup, for
drying.
15.0
CS – Messages displayed
Message
Description
Wait…
HS is being purged
READY
ECU is ready to perform next task
- - - - - - - -
Sample being injected at correct pressure/volume
wait
Sample is being analyzed
Permanent Lockout
ECU has exceeded the period for monitoring by CS read
Cal. Mode
ECU is in calibration mode
Connect Handset
Connect HS to commence calibration
Solution not Ready
ALCOSIM simulator not ready, or not communicating with ECU
- - test(s), - day(s) remaining
Readout of the number of tests and days remaining before fluid change
Change spit trap and solution. Verify 50mg%
Counter has reached the set limit of 25 tests/5 days
Check spit trap. Verify ALCOSIM simulator
temperature and solution 50mg%
Check spit trap for moisture; check ALCOSIM simulator temperature;
check that solution is 50 mg%.
Press Button to Start
Press button to start calibration procedure
Low pressure: Press button to continue
Insufficient air pressure from pump to conduct testing
BAC:
Blood alcohol concentration
Cal. Done
HS calibration procedure completed successfully
VerifyOK
Calibration of HS has been verified
BAC1=
First BAC readout during process
BAC2=
Second BAC readout during process
Old- - - - -
Old calibration factor
New- - - - -
New calibration factor
Last Calibrated
The last date when HS was calibrated.