15
10.0
Automatic disabling
The CS is pre-programmed to disable any HS that does not pass the calibration parameters. Once disabled, the HS cannot be
reactivated. Disabled HSs should be immediately returned to your distributor.
11.0
Resetting the internal counter (changing solution)
After 25 tests or 5 days the
Change spit trap and solution. Verify 50 mg%
message will display.
ATTENTION!
Never use the Alcohol solution beyond 5 days or 25 tests. The HS will be incorrectly calibrated or unnecessarily
disabled.
To reset the internal counter, insert the into the Auto-Cal ECU. Press the
Diagnostic
key and press the
Previous
key
twice until
Bottle #
is displayed on the screen. Verify that the solution value you are using is 50 mg/dL (0.50 g/L). Press the
#
key and enter the bottle number from the new bottle of alcohol solution, then press
OK
. The internal counter is now reset. The
genius can now be disconnected. The bottle number will be stored into the internal memory for later audit.
Every morning prior to turning the ALCOSIM simulator on, check the number of tests available. If only several tests are left you
may wish to change the solution prior to starting calibrations, as warming the solution will take up to 20 minutes.
12.0
Synchronization
Every 60 days, the CS must be connected to ITE for a CS read transaction to download the calibration information for transfer
to the central computer during SYNC operations. If the read procedure is not completed, the CS will display
Permanent
Lockout
until the procedure is complete.
12.1
CS Read Procedure
The CS read procedure is a form of monitoring which is designed to upload the contents of the CS log into the database at
preset intervals.
Open ITE on PC and find the appropriate CS by first clicking on
Find Client Name
button at the top right hand side of the
Clients
screen and choosing the appropriate CS, or by pressing the
Find Connected Client
button.
Start the transaction by clicking on the
New Transaction
button.