А1270 Electro-Magnetic Ultrasonic Thickness Gauge
Operation Manual
31
ACOUSTIC
CONTROL
SYSTEMS
- The current velocity measurement result will be displayed on the screen. After performing all measurements, the velocity
value and the message "Save result?" are displayed on the screen.
The screen layouts (orientation) of the instrument during the calibration procedure are shown subsequently in the Figure 28.
Saving the velocity value obtained during calibration.
Active keys:
F1 (Yes): saving the obtained velocity value for the material existing in the library which shall be selected from the list of the mate-
rials, (Figure 29) or for the new material: select the “New” option, set the name of the material and press the key F1 (Save) (Figure 30).
F3 (No): exits without saving the result.
Figure 29
Figure 30
2