А1270 Electro-Magnetic Ultrasonic Thickness Gauge
Operation Manual
22
ACOUSTIC
CONTROL
SYSTEMS
The additional menu options of the SETUP MEMORY mode and their corresponding parameters (in the metric measurement
system) and functions are listed in the Table 5.
Menu option
(parameter)
Parameter value
Description
Monitor
Off / Outside/ Inside
Off – switching Off of the display.
Inside / Outside – Sets a criterion of the sound alarm respond if the mea-
surement result is within the set limits (inside) or beyond them (outside)
Limit: beginning, mm
from 0 to 150
Sets the lower response limit of the display
Limit: end, mm
from 1.1 to 300
Sets the upper response limit of the display
Clear memory, %
from 0 to 100
Indication of the volume of the used memory. Starts the procedure of
deletion of the measurement results from the memory
Menu option
(parameter)
Parameter value
Description
Scan start, mm
from 0 to 295
Sets the start of the scanning
Scan end, mm
from 5 to 300
Sets the end of the scanning
Gate: beginning, mm
from 0 to 300
Sets the lower boundary of the strobe
Gate: end, mm
from 0 to 300
Sets the upper boundary of the strobe
Amplification, dB
from 0 to 40
Sets the amplification
A-Scan image
Filling / Contour
Selects the visual image of the signal in the А-SCAN mode
Table 5
Table 6
The additional menu options of the SETUP – А-SCAN mode, their corresponding parameters (in the metric measurement
system) and functions are listed in the Table 6.