![ACS A1270 EMAT Operation Manual Download Page 23](http://html1.mh-extra.com/html/acs/a1270-emat/a1270-emat_operation-manual_2841796023.webp)
А1270 Electro-Magnetic Ultrasonic Thickness Gauge
Operation Manual
23
ACOUSTIC
CONTROL
SYSTEMS
Menu option
(parameter)
Parameter value
Description
Range
min-max / 0-max
Selects the image option of vertical scale of the B-Scan range
Scan velocity, mm/s
from 1 to 10
Setts the scanning speed
Table 7
The additional menu options of the SETUP B-SCAN mode, their corresponding parameters (in the metric measurement
system) and functions are listed in the Table 7.
2.3.2.1 The MODE option
Select the measurement mode:
- MEMORY: the measurement results stored in the memory are dis-
played on the screen;
- B-SCAN: the testing object profile is displayed;
- А-SCAN: the signal in form of an A-Scan is displayed on the screen.
Active keys:
F1 (Memory): selects the MEMORY mode;
F2 (B-Scan): selects the B-SCAN mode
F3 (А-Scan): selects the А SCAN mode;
subsequent switching between the modes.
The MODE option is shown on the screen, Figure 19.
Figure 19
2