Special requirements for isolation transformers for power supply in medical locations
.
4 Technical parameters
4.1 Technical parameters of AITR series medical isolation transformer
Refer to Table 2.
Table 2 Technical Parameters of AITR Series of Medical Isolation Transformer
Type
AITR10000S
AITR8000S
AITR6300S
AITR5000S
AITR3150S
Insulation class
H
H
H
H
H
Protection class
IP00
IP00
IP00
IP00
IP00
Power /Voltage /Current
Rated power
10000VA
8000VA
6300VA
5000VA
3150VA
Rated frequency
50-60Hz
50-60Hz
50-60Hz
50-60Hz
50-60Hz
Rated
input voltage
AC230V
AC230V
AC230V
AC230V
AC230V
Rated
input current
45.3A
36A
28.5A
22.5
14.2A
Rated
output voltage
AC230V/115V AC230V/115V AC230V/115V AC230V/115V AC230V/115V
Rated
output current
43.5A
34.7A
27.4A
21.7
13.7A
Inrush current
<12In
<12In
<12In
<12In
<12In
Leakage current
<200μA
<200μA
<200μA
<200μA
<200μA
No load
input current 1.359A
1.08A
0.855A
0.675A
0.426A
No load output voltage 235V±3%
235V±3%
235V±3%
235V±3%
235V±3%
Short circuit voltage <6.9V
<6.9V
<6.9V
<6.9V
<7.5V
General parameters
Fuse wire
80A
63A
50A
35A
25A
Primary
winding
resistance
<55mΩ
<64mΩ
<80mΩ
<131 mΩ
<245mΩ
Secondary
winding
resistance
<45mΩ
<64mΩ
<80mΩ
<116 mΩ
<228mΩ
Iron loss
<150W
<105W
<107W
<77W
<55W
Copper loss
<230W
<200W
<170W
<125W
<120W
Efficiency
>96%
>96%
>96%
>96%
>95%
Maximum
ambient
temperature
<40℃
<40℃
<40℃
<40℃
<40℃
No-load
temperature
rise
<36℃
<33℃
<31℃
<26℃
<22℃
Full load temperature
rise
<65℃
<76℃
<67℃
<62℃
<55℃
Noise grade
<40dB
<40dB
<40dB
<40dB
<40dB