29
byte
CRC check
code
High
byte
53H
Low
byte
3FH
7.1.3 Parameter address table in AIM-M300 series medical intelligent insulation monitoring device
No.
Address
Parameter
Read-write
property
Value range
Data
type
1
0000H
Protecting
passwords
R/W
0001-9999 (Default value is
0001)
Word
2
0001H high
byte
RS485
Communication
address
R/W
1~247 (Default value is 1)
Word
0001H low
byte
RS485
Communication BPS
R/W
1~4
:
4800
、
9600
、
19200bps(Default value is 2)
3
0002H high
byte
CAN
address
R/W
1-110 (Default value is 1)
Word
0002H
low
byte
There is fault
location device or
not
R/W
1:yes,0:no (Default value is
0)
4
0003H high
byte
LCD
contrast ratio R/W
0-63 (Default value is 0)
Word
0003H
low
byte
Backlight timeout R/W
0: Normally open,1-99(Unit is
Min)
5
0004H high
byte
Year
R/W
1-99
Word
0004H
low
byte
Month
R/W
1-12
6
0005H high
byte
Day
R/W
1-31
Word
0005H
low
byte
Week
R/W
1-7