5.3 Wiring method
5.3.1 Wiring mode of AITR series medical isolation transformer
The input terminals at the transformer terminal blocks are labeled with “PM”, in which two
terminals 0 and 230 are connected to the input 220V single-phase AC. The output terminals are labeled
with “SEC”, in which the output voltage of two terminals 0 and 230 is AC 220V and is connected
to external field load. The S terminal is connected to the PE bus bar on the spot (or the equipotential
terminal line). Two ST terminals are temperature sensor interfaces, which are respectively connected
to the No.11 and 12 terminals of AIM-M300 series insulation monitoring device.
Figure 2 AITR series medical isolation transformer terminal blocks diagram