COPYRIGHT 2010 - 2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
46
OMNICYCLE
®
ELITE USER MANUAL
Immunity to Electromagnetic Disturbance (cont.)
For 80MHz and 800MHz, the higher frequency range applies.
These guidelines may not be applicable in every case. Buildings, objects and persons influence the propagation of
electromagnetic waves by absorption and reflection. An exact technical estimate of the field intensity of stationary
transmitters such as base stations of mobile phones and land mobile radio devices, amateur radio stations, AM and
FM radio broadcast and TV stations is not possible.
Directives and Manufacturing Declaration – Immunity to Electromagnetic Disturbance
The Omnicycle
®
Elite is designed for operation in an electromagnetic environment as described below. The customer
or the user should make sure that the unit is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Level of
Conformity
Directive for electromagnetic environment
Conducted HF
3V
eff
3V
eff
Portable and mobile radio equipment should
Emissions
150kHz to 80 MHz
not be used closer to this device than the
IEC 61000-4-6
protective distance recommended. The
equation for the transmission frequency in
Radiated HF
3V/m
3V/m
questions is used to calculate it.
Emissions
80MHz to 2.5GHz
Recommended protective distance:
IEC 61000-4-3
d = [3.5/3] * '1P = 1.17 * '1P
d = [3.5/3] * '1P = 1.17 * '1P
for 80MHz to 800MHz
d = [3.5/3] * '1P = 2.33 * '1P
for 800MHz to 2.5GHz
with (P) as the transmitters power rating in
watts (W) according to the indications of the
transmitter manufacturer and (d) as the
recommended protective distance in meters
(m)
Note: