OMNICYCLE
®
ELITE USER MANUAL
TROUBLESHOOTING
The following table lists machine problem symptoms and possible areas to check for the problem causes. If these
suggested measures do not correct the machine malfunction, call your ACP Customer Support at 800-350-1100.
PROBLEM
REMEDY
Omnicycle
®
Elite has too little
power
•
Shorten crank length
•
Increase motor power one increment
(Note: Be certain not to overstress joints and tendons)
The spasticity detection responds
too frequently
•
Reduce the sensitivity of spasticity detection
•
Shorten crank length
•
Increase motor power one increment
•
Optimize sitting position and distance from the unit
(Note: Be certain to consult with your therapist to make sure your joints and
tendons are not being overstressed)
The resistance is too high for
active exercising
•
In Neuro exercise mode, reduce the resistance
•
In Ortho or Cardio exercise mode, reduce the power setting
•
Increase the crank length
The heart rate sensor does not
work
•
Make sure that the heart rate sensor is properly connected
•
Make sure the heart rate sensor is properly positioned and/or moisten the
heart rate sensor. Reposition the heart rate sensor.
The Omnicycle
®
Elite runs
unevenly
•
Switch ON muscle power assistance
•
Set equal crank length on both sides
•
Check that sitting position is central (relative to spindle)
The Omnicycle
®
Elite does not
start
•
Have you pressed “START” (twice if Omnicycle
®
Elite is in stand-by
mode)? Screen may reset if in standby mode.
•
Check power cord. Unplug the power cord then reconnect.
•
Check power supply fuse
•
Select the correct exerciser (upper or lower body)
•
Unlock the upper extremity exerciser
Exercise session analysis is not
displayed
•
Exercise for at least one (1) minute
Safety voice deactivation does
not respond
•
Increase the microphone sensitivity
Exercise settings have changed
•
Set new basic settings and save
•
Reset basic settings to factory settings
LE / UE exerciser button has no
effect
•
Check connection of upper extremity exerciser
Symmetry bars / road diagram in
Bio-feedback display shows
obviously incorrect data
•
Set equal crank length on both sides
•
Check sitting position is central relative to spindle
•
Are you “pulling” instead of “pushing” with one foot?
•
The Omnicycle
®
Elite needs to be recalibrated
Crank does not stop when using
assisted entry function
•
Press STOP button to stop accidentally started exercising session, then
press and hold START button at least 3 seconds
•
If problem persists, contact the Service Center
Replacing a fuse
1.
Unplug the power cord from the machine
2.
Using a small screwdriver, remove the power supply fuse cover.
3.
Use the screwdriver to remove both fuse holders.
4.
Remove blown fuses.
5.
Insert fuses of the correct rating as per the rating plate in the fuse holders.
6.
Replace fuse holder and fuses.
7.
Replace power supply fuse cover.