25
<
Memory
>
(2/3)
Items
Default Contents
● Mode
:
Normal
:
Normal
:
Basic setting
Auto
:
Automatic measurement, where the following
items are available.
Start
:
Start of power supply ON automatic measuring
system.
(
It records on the memory card simultaneously
)
● Interval
:
Single
:
Measuring interval setting
Single
:
The measurement starts with [Start/Stop] key
and is terminated at Meas.Time selected .
Repeat
:
The measurement starts with [Start/Stop] key
and is repeated in every Meas.Time selected until
[Start/Stop] key is pressed.
【
When Mode
:
Auto is selected, the following items can be specified
】
● I/O
:
OFF
:
External output setting
OFF
:
Default (Data output is disabled).
ON
:
Outputs data for one second when the data
mory mode is active.
● Level
:
65dB
:
65dB
:
Threshold level is registered (when the level
exceeds it, recording starts), within the range
20-130dB at resolution 1dB
● Samp Time
:
Meas.Time
:
Meas.Time
:
sampled at interval equal to Meas.time.
100ms
:
sampled at interval 100ms (0.1s).
200ms
:
sampled at interval 200ms (0.2s).
1s
:
sampled at interval 1s
Meas.Time is time set with [Meas.Time] key (1s
~
-
・・・
).
*
Select 10s or more in L
Atm5
measurement.
● Sta
:
Registers the starting time for recording (YY/MM/DD HH/MM/SS)
(Year/Month/Date, date time/minute/second)
● Stp
:
Registers the stop time for recording (YY/MM/DD HH/MM/SS)
(Year/Month/Date, date time/minute/second)
Fixed to Mease Time, when when RSR card is installed.
Select to
Mode:Auto
Select to
Mode:Start
<Memory>
2/3
Mode
:Normal
Interval :Single
<Memory>
2/3
Mode
:Auto
Interval :Single
I/O :OFF
Level
: 65dB
Samp Time
:MeasTime
Sta:08/10/10 18:16:00
Stp:08/10/12 20:16:00
<Memory>
2/3
Mode
:Start
Interval :Repeat
I/O out :232C
Summary of Contents for 6236