USER MANUAL 16-2000-120
Rev. 02/2016
3. TECHNICAL SPECIFICATIONS
POWER SUPPLY
Power supply voltage
220-240 [V] ± 10% 50/60 [Hz] or
optionally
110-120 [V] ± 10% 50/60 [Hz]
Rated power consumption
300 [VA]
SAFETY CONDITIONS
Electric shock protection:
Class
I
Degree
CF
Degree of casing protection
IP2X
Low-frequency leakage currents
according to EN 60601-1
High-frequency leakage currents
according to EN 60601-2-2
Generator operation frequency
333 [kHz]
Defibrillation impulse resistance
according to EN 60601-1
NEM system
Neutral electrode monitoring.
Maximum resistance at which the unit will operate: 160
[Ohm]
AUTOTEST
Automatic test of the unit and accessories after power
supply is connected. Display of service codes.
OVERLOAD
Overload protection
CUTTING
PURE CUT Pure monopolar cutting
9 levels
120 [W] for 300 [Ω], max 700 [Vp]
BLEND CUT Monopolar cutting with hemostasis
9 levels
120 [W] for 300 [Ω], max 1500 [Vp]
COAGULATION
FORCED COAG Forced monopolar coagulation
9 levels
120 [W] for 300 [Ω], max 1600 [Vp]
SOFT COAG Soft monopolar coagulation
9 levels
120 [W] for 50 [Ω], max 225 [Vp]
PAGE 7