USER MANUAL 16-2000-120
Rev. 02/2016
A periodical inspection is required once a year. The manufacturer only permits the use of units
which an up-to-date inspection has been performed by an authorised service.
The Declaration of Conformity supplied by the manufacturer does not cover devices the
maintenance, services or repairs of which are carried out by unauthorised services.
The manufacturer does not anticipate any calibrations or repairs of the electrosurgical system
to be performed by the user, with the exception of power and mode settings.
Fuses can only be replaced by an authorised service of the manufacturer.
The user is obliged to perform technical inspections recommended by the manufacturer, which should be
performed by a service authorised by the manufacturer. If this condition is fulfilled, the manufacturer remains
responsible for device safety. If the user does not adhere to the manufacturer’s instructions and the required
inspections are not performed, then, according to the laws, the responsibility is transferred to the user.
In order to ensure the correct operation of the device, the installation and staff training should be performed
by an authorised representative of Ackermann. Each participant of such training receives a certificate which
entitles him/her to use Ackermann electrosurgical units. These procedures are obligatory.
More information on the authorised services may be obtained from the manufacturer
In case of a breakdown of the unit, the user must contact the service to obtain the repair
authorisation number. The unit must be delivered to the service (see p.2) after the indicated
number is previously obtained.
UNIT TRANSPORT
Please adhere to standard safety measures when transporting the unit. During transport, the unit must be
protected against mechanical damage and moisture.
If the unit has been transported for a long time, it should be allowed to reach room temperature
before it is started
.
UNIT CHARACTERISTICS
The characteristics presented here can change as our products develop.
FREQUENCY
333 kHz ±10%
MATCHING RESISTANCE
300 Ω
WAVEFORM
SINE WAVE
FREQUENCY
333 kHz ±10%
MAXIMUM REPETITIVE PEAK-TO-PEAK
VOLTAGE
1200 V
MAXIMUM POWER TRANSFER FOR MATCHING
RESISTANCE
120 W ±20%
PAGE 28