
USER MANUAL 16-2000-120
Rev. 02/2016
Check the neutral electrode application and the connected cables every time the patient’s
position has been changed. Protect the neutral electrodes against wetting during the procedure.
9.3.2. REUSABLE NEUTRAL ELECTRODES
When performing surgical procedures:
- requiring high power settings (e.g. TURP);
- posing a risk of flooding the neutral electrode with liquids;
- in which the staff is not able to control the neutral electrode application;
- on children
THE USE OF REUSABLE NEUTRAL ELECTRODES IS NOT ALLOWED
When using one-piece silicone electrodes, the surgical team is fully responsible for the correct
application of an electrode. Therefore, take special care to correctly attach the neutral electrode
to avoid burns at its application site during the procedure. The application of a one-piece
neutral electrode should be monitored throughout the duration of the procedure.
Before applying a neutral electrode, read the instructions for use supplied by its manufacturer.
A reusable one-piece neutral (silicone) electrode does not enable the application to be
monitored by the unit, i.e. to control its adhesion to the patient’s body. Only a correct electrode
connection to the unit is monitored.
The neutral electrode must not be wetted or wrapped with anything. Do not spread any
additional conductive gel on the surface of the neutral electrode. When disconnecting the
neutral electrode, never do it by pulling the cable. Under no circumstances try to repair the
neutral electrode by yourself.
Examine the condition of the neutral electrode and the connection cable before application. Do
not use them in case of visible surface defects on the electrode surface or damaged insulation.
Attach the reusable silicone electrode in such a way as to prevent it from moving spontaneously
– use a special tape for fixing a neutral electrode. Prevent any fluid intrusion between the
electrode and the patient’s body. Reusable neutral electrodes should be disinfected before use
(see section 12 “Unit and accessories maintenance”).
Remember that a silicone electrode loses in time its conductive properties as active substances
are washed out from the silicone. Such an electrode increases the risk of patient burns.
Therefore, reusable electrodes should be subjected to regular inspections, along with units.
Read the manufacturer’s instructions for use before applying a neutral electrode.
ALWAYS comply with the manufacturer’s instructions on the package of the neutral
electrode.
PAGE 18
OF 39