USER MANUAL 16-2000-120
Rev. 02/2016
Neutral electrodes other than those mentioned above may not ensure proper cooperation with the NEM
neutral electrode safety system.
The manufacturer is not responsible for the use of Ackermann HF units with neutral electrodes other than
those mentioned above, or for any incidents resulting from such use.
If the neutral electrode is correctly connected, then a green diode is on at the neutral electrode indicator. If the
neutral electrode is not connected or the electrode with adhesion monitoring is incorrectly applied, after the
unit has been started a green diode is on, a modulated acoustic signal can be heard and the message “EL
NEU-[]” is displayed. It is impossible to activate the unit in the monopolar mode unless the neutral electrode is
correctly connected. If such a message appears during the operation of the unit, stop the operation and check
the connection of the neutral electrode.
An important advantage of the neutral electrode monitoring system is that the monitoring is performed on a
continuous basis, also during the operation of the unit. The neutral control monitoring system does not affect
the operation in the bipolar mode
.
Figure 9
. The message that the neutral electrode is not connected correctly.
9.3.1. DISPOSABLE NEUTRAL ELECTRODES
The neutral electrode must not be modified in any way. Once applied, the electrode should not
be transferred to another place. Never use electrodes after their expiry date. Do not use force
to remove the electrode. It should be unstuck carefully.
Before applying a disposable neutral electrode, dry the application site very carefully. When
alcohol-based disinfectants have been used wait for a while to let the alcohol evaporate. When
using disposable electrodes, always check their expiry date. A disposable electrode can only be
used once.
The neutral electrodes are supplied in closed sachets. After a package has been opened an
electrode must be used within 15 days. After that time, the conductive substance dries out and
does not ensure sufficient conduction. Disposable electrodes should be applied carefully. If it is
necessary to apply the electrode in a different place, use a new electrode.
PAGE 17