MR9610 Microplate Reader Operation Manual Chapter 4 Installation
─
5
─
Chapter 4 Installation
1. Initial check
Each MR9610 is thoroughly tested before shipping, but please check it again when you
receive the instrument and contact your local distributor or Accuris Instruments if:
• The outer package is damaged
• The outer package has any obvious moisture stains
• The outer package has marks of impact
After opening, please check the instrument and box contents.
Confirm that all ordered accessories have been included.
Check the instrument’s appearance for any damage.
2. Installation location and environment
Installation location: locate the instrument on a flat, dry, clean and stable surface, allow
enough room on the front side so there is space for plate holder drawer to move in and out,
and also keep 15cm / 6 inches of space behind and on the sides for any wires and air flow.
Environment:
a. The environment for installation should not be excessively humid or dusty.
b. Temperature should be within the range of +10°C ~ +40°C.
c. Relative humidity should be within the range of 30% ~ 80% to avoid condensation.
Note: KEEP THE INSTRUMENT AWAY FROM CORROSIVE GASSES AND
LIQUIDS.
3. Installation steps
①
Place the instrument on a stable and level surface.
Note: DO NOT loosen any screws or parts without first consulting with the
Accuris Instruments Service Department or your authorized distributor. Doing
so can cause damage and void the warranty.
②
Connect the instrument to an appropriate power outlet using the provided power
cord.