82
Chapter 6: Technical Information
Product Limitations
Please read the literature packaged with your test drum to find the latest information on product specifications
and limitations.
Specifications
Display
Organic Light Emitting Diode (OLED) display
Automatic power-off
After 60 seconds / 5 minutes (depending on testing status)
Power supply
AAA alkaline batteries (do not use rechargeable)
Battery life
1,000
tests
Interface
Infrared
interface
LED/IRED
Class
I
Meter type
Reflectance
photometric
Measuring range
10 – 600 mg/dL
Sample size
Refer to the test drum package insert
Measuring time
5
seconds
System operating conditions
50 °F to 104 °F (10 °C to 40 °C)
Meter storage conditions
With batteries and test drum: 36 °F to 86 °F (2 °C to 30 °C)
Test drum storage conditions
36 °F to 86 °F (2 °C to 30 °C)
Relative humidity operating range
15% - 85% relative humidity
Memory capacity
500 blood glucose values with time and date
Dimensions
4.9 x 2.5 x 1.3 inches (121 x 64 x 32 mm) with lancing device
4.8 x 2.2 x 1.3 inches (121 x 57 x 32 mm) without lancing device
Weight
Approximately 5.2 ounces (147 g) with lancing device, batteries and test strip drum
Lancing device depths
0.80, 0.95, 1.10, 1.25, 1.40, 1.55, 1.70, 1.85, 2.00, 2.15, 2.30 mm