![background image](http://html.mh-extra.com/html/accu-chek/aviva-expert/aviva-expert_standard-owners-booklet_455904078.webp)
72
j
The label on your test
strip container shows the
acceptable ranges for both
the Level 1 and the Level 2
control solutions. The test
result you get should be inside
this range. ensure you select
and compare the test result to
the correct level of control. For
the level selected, the meter
displays whether the result
is within or outside of the
acceptable range.
Example
Range (mg/dL)
Level 1 25-55
Level 2 255-345
j
When the control test result
is inside the range on the test
strip container, the test strips
and the meter are working
properly.
10:02 am
2 Feb 08
Control Result
Menu
mg/dL
Within range
50
Level 1
If the control test result is within
the acceptable range, “Within
range” appears on the detailed
control Result screen.
Example Control Results
45702_std.indb 72
10/29/09 1:42:13 PM
Summary of Contents for Aviva Expert
Page 12: ...6 45702_std indb 6 10 29 09 1 41 47 PM ...
Page 66: ...60 45702_std indb 60 10 29 09 1 42 08 PM ...
Page 224: ...218 45702_std indb 218 10 29 09 1 42 51 PM ...
Page 248: ...242 45702_std indb 242 10 29 09 1 42 53 PM ...
Page 268: ...262 45702_std indb 262 10 29 09 1 42 53 PM ...
Page 278: ...272 45702_std indb 272 10 29 09 1 42 54 PM ...
Page 292: ...286 45702_std indb 286 10 29 09 1 42 55 PM ...