208
Alarm Clock Reminders for bG Test, Injection and Other
• At the scheduled time, the meter turns on and displays the reminder (if a test strip has not been inserted).
• When the meter displays an Alarm Clock reminder and you do not press any meter buttons, after 30
seconds the meter turns off and displays the reminder again in 2 minutes. After the fourth time the meter
displays the reminder, then the meter dismisses the reminder.
• bG Test Alarm Clock Reminder
• If the meter is already on when the reminder is scheduled and no blood sugar test was performed, the
reminder is displayed when the meter powers down.
• Dismiss all active bG Test reminders by pressing
.
.
• Snooze the reminder for 15 minutes either by pressing
,
or by inserting a test strip.
• When you perform a blood sugar test, the meter dismisses any bG Test reminders that have been
snoozed or that are pending within the next 30 minutes.
45702_std.indb 208
10/29/09 1:42:50 PM