![background image](http://html.mh-extra.com/html/accu-chek/aviva-expert/aviva-expert_standard-owners-booklet_455904254.webp)
248
9.4 Product Safety Information
Electromagnetic Compatibility
This meter meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A. The chosen
basis for electrostatic discharge immunity testing was basic standard Iec 61000-4-2. In addition, it meets
the electromagnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus low.
Interference from other electrically driven equipment is not to be anticipated.
Performance Analysis
The performance data for the Accu-chek Aviva
Expert system (ACCU‑CHEK Aviva
expert meter with
ACCU‑CHEK Aviva test strips) were obtained using capillary blood from diabetic patients (method comparison,
accuracy), venous blood (repeatability), and control solution (reproducibility). The system is calibrated with
venous blood containing various levels of sugar. The reference values are obtained using the hexokinase
method. For method comparison, the results were compared with results obtained using the hexokinase m
ethod with deproteinization (automatic analyzer). The hexokinase method is traceable to an NIST standard.
45702_std.indb 248
10/29/09 1:42:53 PM
Summary of Contents for Aviva Expert
Page 12: ...6 45702_std indb 6 10 29 09 1 41 47 PM ...
Page 66: ...60 45702_std indb 60 10 29 09 1 42 08 PM ...
Page 224: ...218 45702_std indb 218 10 29 09 1 42 51 PM ...
Page 248: ...242 45702_std indb 242 10 29 09 1 42 53 PM ...
Page 268: ...262 45702_std indb 262 10 29 09 1 42 53 PM ...
Page 278: ...272 45702_std indb 272 10 29 09 1 42 54 PM ...
Page 292: ...286 45702_std indb 286 10 29 09 1 42 55 PM ...